Baxter Corporation’s MiniCap Extended Life Peritoneal Dialysis Transfer Sets: Certain cleaning products, including hand sanitizers and solvents, may damage sets

Class not statedclass recorded on the notice
2022-02-16last updated
Medical devicescategory
Nomarked archived

What the notice records

Every field on this notice, as published
ProductMiniCap Extended Life Peritoneal Dialysis Transfer Sets manufactured by Baxter Corporation (see list of affected products)
IssueMedical devices
What you should doAvoid direct contact between the transfer sets and the following cleaning products:hand sanitizers and cleaning products that contain hydrogen peroxide, bleach, alcohol or antiseptic agents; andsolvents intended to remove adhesive residue, such as those containingacetone, toluene, xylene or cyclohexanone.If your transfer set is damaged (e.g., cracked or leaking), contact your health care provider immediately for a replacement. Consult your health care provider with any questions or concerns about your transfer set.Seek immediate medical attention if you experience symptoms of peritonitis. Symptoms include nausea, vomiting, and abdominal pain or tenderness. Use proper cleaning practices to clean the device, as outlined in the Instructions for Use or as directed by a health care professional.You may also contact Baxter Corporation at 1-866-968-7477 if you have questions about this issue or about your transfer set.Report any health product-related complaints to Health Canada.
CategoryMedical devices
Publishing bodyCommunications and Public Affairs Branch
ClassClass not stated
Last updated2022-02-16

Read the official notice on recalls-rappels.canada.ca

Everything in Medical devices · Everything from Communications and Public Affairs Branch

Read this correctly

This page reproduces the fields published on the notice and links to the notice itself. It is not a rating and not a statement about a product's current condition, availability or safety.

This notice carries no class in the published data. A blank class means the field is blank — it is not a statement that the product was low risk.

Privacy

This site publishes records that governments already publish. It sets no cookie and makes no request to Google unless you accept one. Read more below.

About this record

Every recall notice published by Health Canada: the product, the issue recorded against it, the class assigned, and the official notice itself. A record of what was published, on the date it was published.

Source: Health Canada — Recalls and safety alerts (open data) — Open Government Licence – Canada. Data refreshes daily. Every page links the row it came from.

What this site is not. It is a copy of an official record. It is not a judgement about any business, building or person, and it is not advice. If the record is wrong, the source above is where it is corrected.